Reference Gauge
• Excellent vibration/shock resistance due to the design of the spindle guide. • Sliding durability improved to remain serviceable for at least 15 million cycles (inhouse testing). • Shock resistance, 100 G/11 ms (IEC 60068-2-27) Linear Gauge ลิเนียร์เกจ
• The LGF-Z series, which is equipped with a reference point mark on the linear encoder (refer to page G-9), and includes a 0.1 µm resolution type (refer to page G-8) is also available.
542-173, -163
542-171, -161
542-172, -162
Dimensions
Unit : mm
542-171, -161
542-171, -161
542-171, -161
Specifications Order No. Measuring range
542-171
542-161
542-172
542-162
542-173
542-163
10 mm
25 mm
50 mm
Resolution
0.5 µm
1 µm
0.5 µm
1 µm
0.5 µm
1 µm
Measuring accuracy (20 °C) L=arbitrary measuring length (mm)
(1.5+L/50) µm
Quantizing error
±1 count
Contact point downwards Contact point horizontal Contact point upwards
1.2 N or less 1.1 N or less 1.0 N or less
4.6 N or less 4.3 N or less 4.0 N or less
5.7 N or less 5.3 N or less 4.9 N or less
Measuring force
Position detection method
Photoelectric linear encoder
Response speed
1500 mm/s
90° phase difference, differential square wave (RS-422A equivalent), minimum edge intervals: 1000 ns for 5 µm model, 500 ns for 1 µm model, 250 ns for 0.5 µm model
Output
Measuring force* 1
2 µm
4 µm
2 µm
4 µm
2 µm
4 µm
Mass
Approx. 260 g
Approx. 300 g
Approx. 400 g
Contact point
ø3 mm carbide tipped (fixing screw: M2.5 (P=0.45)×5), standard contact point : 901312
Stem
ø8 mm
ø15 mm
Bearing
Linear ball type
Output cable length
2 m (directly from casing)
Connector
Plug : RM12BPE-6PH (HIROSE), Compatible receptacle : RM12BRD-6S (HIROSE)
Operating temperature (humidity) ranges Storage temperature (humidity) ranges
0 to 40 °C (RH 20 to 80 %, non-condensing) –10 to 60 °C (RH 20 to 80 %, non-condensing)