Reference Gauge
Measuring Microscope กล้องจุลทรรศน์แบบวัดค่า
• Observation with a clear and flareless erect image along with a wide field of view. • Measuring accuracy is the highest in its class (and conforms to JIS B 7153) • ML series, high-NA objectives that are specially designed for the MF series (long working distance type). • Illumination unit (reflected/transmitted) can be selected from a high-intensity LED or halogen bulb. • Variable aperture diaphragm (reflected / transmitted) allows observation measurement while suppressing light diffraction. • Variety of standardized stages in sizes up to 400×200mm. • Quick-release mechanism useful for moving the stage quickly when measuring workpieces that are large in size or quantity. • Coarse/fine feed handles are as standard on both sides allowing precise focus and observation measurement regardless of handedness. • High-magnification eyepiece observation up to 2000X. • Standard measuring microscope that has a wide variety of optional accessories including a Vision Unit and various digital CCD cameras.
MF-B2017D • The binocular tube (eyepiece) and illumination unit are optional accessories.
Specifications
Model No. Order No. Model No. Order No.
MF-A1010D 176-861* 1 MF-B1010D 176-866* 1
MF-A2010D 176-862* 1 MF-B2010D 176-867* 1
MF-A2017D 176-863* 1 MF-B2017D 176-868* 1
MF-A3017D 176-864* 1 MF-B3017D 176-869* 1
MF-A4020D 176-865* 1 MF-B4020D 176-870* 1
Without Z-axis scale
With Z-axis scale
Observation image
BF (Bright field)/Erect image
10X (eyepiece field number: 24), 15X, 20X Note: Monocular - one 10X eyepiece provided as standard; Binocular - two 10X eyepieces provided as standard Transmitted illumination: Telecentric system, Built-in aperture diaphragm, White LED light source, stepless light intensity control, With cooling fan Reflected illumination: Koehler illumination, Variable aperture diaphragm mechanism, White LED light source, stepless light intensity control Control unit: Power ON/OFF switch (main switch), 100 - 240V AC power input connector Transmitted illumination: Telecentric system, Built-in aperture diaphragm, Halogen bulb (12V, 50W), stepless light intensity control, With cooling fan Reflected illumination: Koehler illumination, Variable aperture diaphragm mechanism, Halogen bulb (12V, 50W), stepless light intensity control, With cooling fan Control unit: Power ON/OFF switch (main switch), 100 - 240V AC power input connector ML objective 3X (provided as standard), 1X, 5X, 10X, 20X, 50X, 100X
Eyepiece (optional) Adjustable diopter
Objective (optional)
LED illumination unit
Illumination unit (One of the two options must be selected.)
Halogen illumination unit
Measurement range
100×100mm
200×100mm
200×170mm
300×170mm
400×200mm
Quick-release mechanism Zero-set button
Stage
Provided as standard for the X and Y axes
Provided as standard for the X and Y axes (and for the Z axis only for the MF-B type)
Max. workpiece height 150mm
220mm
Z axis
Feed mechanism
Coaxial coarse and fine feed, handles on both sides (coarse: 30mm/rotation, fine: 0.2mm/rotation)
Measuring accuracy *2 (X and Y axes, when not loaded)
(2.2+0.02L)μm L: measuring length(mm)
Minimum reading
1/0.5/0.1μm .0001”/.00005”/.00001” switchable X and Y (or X, Y, and Z only for the MF-B type)
Digital display
Display axes
Functions Zero-setting, direction switching, RS232C output, USB output (specific to QSPAK) *1: The following suffixes are added to the order No.to specify the User Manual's language: -10 for English; -11 for Simplified Chinese; No suffix for Japanese. *2: Measuring method complies with JIS B7153.
Standard: Halogen bulb (12V, 50W) No.513667 Bulb life: 1,100 hours
Bulb replacement for transmitted/reflected illumination
JSR Entech Catalogue 202
jet@jet.co.th
jetco@jet.co.th
Powered by FlippingBook