Reference Gauge
QI- Non-contact 2D Vision Measuring System Quick Image Non-contact 2D Vision Measuring System
• A single click enables multiple measurements in one display. A batch measurement can be applied to multiple workpieces in the display after executing a pattern search based on the workpiece position. • This series is equipped with a 3-megapixel color camera. Even with low magnification, high repeatability can be obtained. • The choice of five stage sizes makes it easy to choose a machine to suit the user's application. • The video window automatically displays the measurement data, which enables quick verification.
• This series of manual 2D vision measuring machines offers high-efficiency measurement by employing a telecentric optical system that has a deep focal depth and a wide view monitor. • The stitching function enables the entire display of a large workpiece so that highly accurate and speedy measurement can be performed. • A model equipped with a motorized stage has been added to the series to offer easy and comfortable stage operation.
A motorized stage
QI-C2017D
Specifications
Manual stage model
Motorized stage model
0.2X 0.5X
QI-A1010D QI-A2010D QI-A2017D QI-A3017D QI-A4020D QI-C2010D QI-C2017D QI-C3017D 100×100 mm 200×100 mm 200×170 mm 300×170 mm 400×200 mm 200×100 mm 200×170 mm 300×170 mm 170×170 mm 242×140 mm 260×230 mm 360×230 mm 440×232 mm 242×140 mm 260×230 mm 360×230 mm QI-B1010D QI-B2010D QI-B2017D QI-B3017D QI-B4020D
Order No.
Measuring range (X×Y) Effective stage glass size Maximum stage loading *
Approx.10 kg
Approx.20 kg
Approx.15 kg Approx.10 kg
Approx.20 kg
Main unit mass
Approx.65 kg Approx.69 kg Approx.150 kg Approx.158 kg Approx.164 kg Approx.72 kg Approx.153 kg Approx.161 kg
* Does not include extremely offset or concentrated loads Order No.
QI-A/QI-C 32×24 mm
QI-B
View field
12.8×9.6 mm
Measurement mode Travel range (Z axis)
High resolution mode / Normal mode* 4
100 mm
High resolution mode
±2 μm ±4 μm ±1 μm ±2 μm
±1.5 μm ±3 μm ±0.7 μm ±1 μm
Measurement accuracy within the screen* 1 Repeatability within the screen (±2 σ )* 2
Normal mode
Measuring accuracy
High resolution mode
Normal mode
±(3.5 + 0.02L) μm L=arbitrary measuring length (mm)
Measurement accuracy (E1xy)* 1
7.6X 0.2X
18.9X 0.5X
Monitor magnification* 3
Magnification (Telecentric Optical System)
±0.6 μm ±11 μm
±0.6 μm ±1.8 μm
Optical system
Depth of focus Working distance
90 mm
Camera
3 megapixel, CMOS color camera (1/2 in) Green LED telecentric illumination
Transmitted light Co-axial light
Illumination
White LED
Ring light
4-quadrant white LED AC100 to 240 V 50/60 Hz
Power supply
Accuracy guaranteed temperature
20±1 ºC
*1 Inspected to Mitutoyo standards by focus point position. *2 The measuring accuracy is guaranteed to be accurate within the depth of focus. *3 For 1X digital zoom (when using a 22-inch-wide monitor) *4 Patent registered (Japan)